Check the storage temperature of the AED. The device should be stored in an environment between 0°C and 50°C (32°F and 122°F).
If the device is outside this range, move it to a location within the recommended temperature range.
Allow at least 30 minutes for the device to reach normal operating temperature.
Open the lid of the AED.
Wait for three seconds to pass on the timer, then press and hold both buttons on the AED display panel until the voice prompts “diagnostic mode”. The display panel shows TEMP TOO HOT or TEMP TOO COLD and ADJUST STORAGE TEMP.
Press and hold the button under the word CLEAR until the error clears from the display.
Close the lid. The Rescue Ready indicator changes to green in a few short seconds.
Defib. Lifespan
G5 - 8 year warranty
ipad - 7 year warranty, extendable to 10 years, on registration
Batteries
G3 - 4 years
G5 - 4 years, 500 shocks
ipad - 5 years, 200 shocks @150 joules
Pads
Ipad - 2 year lifespan
G5 Paed. - 12-24month lifespan. For <8yo or <25kg. When connected, the energy levels are reduces